Spectrophotometric determination of palladium with sulfochlorophenolazorhodanine by flow injection

dc.contributor.authorShiundu, Paul M
dc.contributor.authorWentzell, Peter D
dc.contributor.authorWade, Adrian P
dc.date.accessioned2015-05-21T15:25:26Z
dc.date.available2015-05-21T15:25:26Z
dc.date.issued1990
dc.description.abstractSulfochlorophenolazorhodanine (as its sodium salt) has been used in the automated development of a sensitive flow-injection procedure for the spectrophotometric determination of palladium. The resulting method has high sample throughput, good precision, and low consumption of both sample and reagents. The optimum pH for the reaction is 5.0 and the response is constant at pH between 4.7 and 5.3. The sensitivity (calibration slope) of the procedure is 4.4 × 103 l./mole. The linear dynamic range is 0.045–30.0 μg/ml. The sample throughput is at least 120/hr. An automated procedure for optimization of analytical variables is described and a two-variable response surface for the system is given. Interference studies on 19 metal ions show that the method has good selectivity.en_US
dc.identifier.citationTalanta Volume 37, Issue 3, March 1990, Pages 329–336en_US
dc.identifier.urihttp://www.sciencedirect.com/science/article/pii/003991409080062K
dc.identifier.urihttp://hdl.handle.net/123456789/1032
dc.language.isoenen_US
dc.titleSpectrophotometric determination of palladium with sulfochlorophenolazorhodanine by flow injectionen_US