DETERMINING TRANSIT DOSE FOR THE BEBIG SAGINOVA® HDR Ir-192 STEPPING SOURCE BRACHYTHERAPY UNIT
Loading...
Date
2023-11-01
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
Technical University of Kenya
Abstract
The Bebig SagiNova® HDR brachytherapy treatment unit implements the TG-43 formalism in calculating radiation dose, which does not account for transit doses during treatment. It is unclear how significant the transit dose is and depending on its magnitude, it may impact the clinical outcome. The goal of this study is to determine the transit dose component for the Bebig SagiNova® HDR brachytherapy unit with an Ir-192 stepping source. The well-type chamber measurement technique was used to measure charge collected as the Ir 192 source moved from the afterloader. The charge measurements were collected for different source configurations and analyzed using two techniques; the multiple exposure method and the graphical method to determine effective transit time. The overall effective transit time was quantified as the source moved to its first dwell position (entry time), between activated dwell points (interdwell time) and during retraction out of the applicator back into the afterloader (exit time). The effective transit time of 2.02s was obtained with the multiple exposure method as well as the graphical solution method. The effective transit time was not influenced by the analysis technique. The overall effective transit time for the unit was determined to be about 13.80s and 13.99s using the multiple exposure method and the graphical solution respectively. The significance of the amount of dose during transit is not clear, as it depends on the activity and configuration of the source, prescribed dose and the quantity of treatment fractions used. It is necessary to determine and document transit time and doses for assessing their significance on the delivered dose to help improve the overall efficiency of brachytherapy and patient care